How to Solve Water System Fault When the Temperature and Humidity Test Chamber in Operation

Oct 12, 2019

Have you encountered the water system fault when operation the temperature and humidity test chamber

Today, let me talk about common problems and solutions for the equipment. We specifically divide water system fault as water leakage and water shortage. Firstly, we look at water leakage.


In the case of water leakage in the temperature and humidity test chamber, water leakage is mainly caused by wet ball grooves and water circulation leakage.

  • The wet-bulb leaks. The wet ball slot is located below the wet ball cloth. For this case, it is common that the ground is uneven, resulting in sloped water leakage in the wet ball groove This requires the replacement of the flat position of the ground. Of course, this situation should also be checked from the water inlet of the humidity test chamber. There is water inlet valve involved in the water supply to the wet ball tank. It is possible that the inlet valve is opened relatively large. You can try to adjust the close the inlet valve of the test equipment.


  •  The waterway is leaking. This is caused by the rupture of the water tank and the water pipeline, and the repair of the gap weld or the related joint operation can be directly solved.



If there is a water shortage problem in the temperature and humidity test chamber, it can be divided into water shortage of the wet ball cloth and water shortage of the water circulation system.


  •  Wet ball cloth lacks water.

    Assuming that the humidity is rising, even when it reaches 98% RH, the wet ball cloth is mostly lacking in water. It is possible that impurities in the water precipitate on the wet cloth and lost the ability to absorb water. So at this time, check if the wet cloth is too dirty.

    If so, replace it with a wet cloth. If the wet ball cloth is new, there is still such a situation. Look at the wet bulb of the high and low temperature damp heat test chamber for water. If not, it is a problem with the water supply circulation system. This detects the pump and the failure of the water level float described above. 

    There is also has a point to note: Is the lower part of the wet cloth soaked in the water of the wet cloth sink?


  • The water system is short of water. 

    In the most cause, the water shortage in the waterway is caused by water shortage in the water tank. About this situation, the purified water is added to the water tank until the water level meter on the lower right side of the water tank reaches the entire display level of about 80%.

TH-100


All of the above is to share with you all the solutions to the problem of water system fault in climate machine.

We have more than 19 types of temperature and humidity machines

Welcome to consult anything you are interested about our chamber!